Publications
> 'Separation efficiency of the MASHA facility for short-lived mercury isotopes'
Separation efficiency of the MASHA facility for short-lived mercury isotopes
Author
Year
2014
Scientific journal
Hyperfine Interactions, 227 (1-3), 209-221
Web
Abstract
The mass-separator MASHA built to identify Super Heavy Elements by their mass-to-charge ratios is described. The results of the off- and on-line measurements of its separation efficiency are presented. In the former case four calibrated leaks of noble gases were used. In the latter the efficiency was measured via 284 MeV 40 Ar beam and with using the hot catcher. The ECR ion source was used in both cases. The α -radioactive isotopes of mercury produced in the complete fusion reaction 40 Ar+ 144 Sm →184−xn Hg+xn were detected at the mass-separator focal plane. The half-lives and the separation efficiency for the short-lived mercury isotopes were measured. Potentialities of the MEDIPIX detector system have been demonstrated for future use at the mass-separator MASHA.
Cite article as:
A. Rodin, J. Klíman, L. Krupa, I. Siváček, S. Pospíšil, . et al., "Separation efficiency of the MASHA facility for short-lived mercury isotopes", Hyperfine Interactions, 227 (1-3), 209-221 (2014)